EchoBridgeMD Pearl #5: EPSS -- A Rapid Bedside Clue to Reduced Left Ventricular Systolic Function
- Stephen Alerhand
- Jul 9
- 1 min read
Updated: Jul 31

When evaluating a patient with suspected left ventricular (LV) systolic dysfunction, E-point septal separation (EPSS) provides a rapid, quantitative estimate of left ventricular ejection fraction.
In the parasternal long-axis view using M-mode, EPSS is measured as the shortest distance between the anterior mitral valve leaflet at its maximal early diastolic excursion (E-point) and the interventricular septum.
As LV systolic function declines, the mitral valve opens less widely while the LV progressively dilates, increasing the EPSS distance.
An EPSS > 7–8 mm is generally considered abnormal and should raise suspicion for reduced LV ejection fraction.
Reference:
📖 Alerhand S, Adrian RJ, Taylor LA. Cardiac Point-of-Care Ultrasound: An Emergency Medicine Review. Emerg Med Clin North Am. 2024.